Sign In | Join Free | My himfr.com |
|
Ask Lasest Price | |
Brand Name : | Phymetrix |
Model Number : | DPT-820 |
Payment Terms : | T/T |
Delivery Time : | 5-8 Working Days |
Loop powered Moisture Analyzer , Dew Point Transmitter , Range -110℃ dp to +20℃ dp
DPT 820
Sensor Operating Principle
The nanopore sensor utilized in this analyzer measures the amount
of water vapor molecules that
have entered its pores.This amount is directly proportional to the
partial water vapor pressure in
the gas surrounding the sensor,and the partial water vapor pressure
is dependant on the water
vapor content of the gas and the total pressure of the gas,Thus to
compute the water vapor content
of the gas under measurement the analyzer has to “know”the total
pressure of the gas.Since there
is no built-in pressure sensor,the user should enter the total
pressure into the analyzer,this can be
easily done by entering atmospheric pressure(the analyzer default
setting)and measuring at
atmosphericpressurebyallowingthesample celloutlet(exhaust)port
toventwithoutrestrictions
Display | DP (℃), PPMV, RH %, either one |
Power supply | DC12V - DC30V |
Output | 4-20 mA |
Sensor | Resistance capacity |
Alarm | Colors: Red: High alarm Pink: low alarm White: test in range |
Range | 110 ℃ ~ + 20 ℃ |
Accuracy | 2 ℃(temperature corrected) |
Repeatability | 0.8 ℃ |
Response time | T95 = 3 min |
Sampe flow | > 1LPM |
Dimensions | 2. include sample chamber: 4.331″*2.362″*4.760″(L*W*H) (110 mm* 60mm * 120.9 mm) |
Weight | Transmitter$sensor:14.11 oz.(0.4Kg) Sample chamber:29.98 oz.(0.85Kg) |
Explosion-proof grade | IP67 (Designed) |
Inlet&Outlet | 1/8 NPT 316L-SS |